UV-VIS SPECTROPHOTOMETER
  Ideal for Biochemical,Analytical, Agricultural
and Medical Research


Split-beam Optics, Dual detectors
1.0nm Bandwidth
Sipper System
PC interface Windows Program
 
 
 
SP-3000 plus implements
high precision/resolution with high precision Sine Driver Monochromator and stable voltage and signal
process by adopting Noise Reduction Circuit
SP-3000 plus can be operated by standalone mode or PC control mode.
The Software included in SP-3000plus can be used in general quality and quantity analysis
Furthermore, OptiView (PC interface windows program) provides a more convenience for user in the
various application fileds
Specificity of SP-3000 plus
ABM (Auto Blank Measuring) Method
Before measuring each sample, automatically measuring blank so error caused by long time
measurement reduce to a minimum.
AGC (Automatic Gain Control) electric circuit adoption.
The intensity of light source is not constant for all measurement wavelength range, so the
sensitivity is varied according to the measuring wavelength. To solve this problem, SP-3000 plus automatically divides the gain of photodiode preamplifier into optimal ranges
LC Mirror Changing System
Lamp changing mirror is directly connected with LC motor so backrush of belt type is reduced
to a minimum. Because of automatically select the maximum intensity of radiation every turn on, SP-3000 plus has good intensity of radiation and precision of wavelength
  A large number/quantity sample measurement at a time

SP-3000plus basically offers Automatic cell holder (eight position cell, rotary type). 8-cell holder of
Rotary type adopted precision light path design and micro stepping control system can measure sample more easily, conveniently, quickly also can measure continuously many samples with
a sipper.
  Easy Interface

SP-3000plus basically offers Standard curve, Time scan, Survey scan, ABS Ratio (DNA/Protein) as
well as ABS/ %T/Conc. for wide application.
Also OptiView (PC interface Windows program) enlarge the application fields of SP3000plus by merging
graph, transferring data, applying to specific conversion formula.
  Useful Utilities

SP-3000plus can check light source condition, correct wavelength and select one of the 13
concentration unit. OptiView can automatically convert diluted concentration value into original value
and save graph/data to Excel or imaging file. Also can print various reporting form.

Self Correction Mode
Self correction for wavelength and cell holder position. Measured value for wavelength and cell holder
position is compared with input correction data and then saved automatically.
Self diagnosis

By checking lamp and wavelength correction condition, get a more reliable data.
By testing intensity of radiation at specific wavelength in ultraviolet and visible range, can confirm its
condition is good or not.
Lamp Change Mode
Setup the wavelength for changing point between deuterium lamp and tungsten-halogen lamp in range
in range 350nm to 470nm for measuring with the optimum intensity of radiation.
ABS/% T/CONC

Absorbance/Transmittance/Concentration Measurement mode. Measuring one to ten samples for
multi wavelength at a time. Saving measured method (Max 10).
Concentration calculation with a standard curve or factor (Y=A*X+B; slope, B; intercept)
Survey Scan

Measure continuously absorbance/transmittance in 190nm~1100nm and confirm the data by Zoom in/
out graphs in specific range. Search for Peaks and Valleys values automatically.
Standard Curve

Can measure by 7 Standard samples at a time and select one of 13 kinds concentration unit, Linear(N),
Linear(O), Segmented Standard curve
Types. Also can Make Standard Curve by inputting the maximum
30 of Concentration and Absorbance.
ABS Ratio

Can measure DNA/Protein by Numerator, Denominator, Reference measurement and factor.
260, 280, 230nm and correction wavelength(320nm) are setup basically to measure more easily.
Simple Kinetics

Measure continuously absorbance/transmittance variation at one wavelength during a certain period.
Measuring interval is over 0.1 second and total measuring time is 99 hours in a maximum
 (OPTIONAL)
Can check condition of SP-3000
plus with OptiView, use SP-3000
plus of optimal condition by
checking cell holder position,
intensity of radiation, stray light,
photometric accuracy, wavelength
accuracy.
By start-up function of OptiView, can measure directly on executing
OptiView.
Measured data is saved by mode, so can find the file easily.
Measurement Contents
1.ABS/%T/Cone
* Time, cell No, wavelength, ABS/%T/Conc indication
* Multi wavelength measurement: Max 10
* Function for selecting sample: Max 8
* Data saving: table or various report form
* Input measuring method by 10 methods.
* Automatic quantitative analysis: using factor or standard curve.
Select one of 13 concentration unit.
* Memo: sample name or etc.
* Data transfer: Excel or text file
2. Survey Scan
* Wavelength setup range: 190nm~1100nm
* Select cell No. : B, 1,2,3,4,5,6,7
* Measuring progress indication with(%)
* Graph Zoom In/Out
* Graph setting: by wavelength or section
* Clip board function: scan data overlapping
* Data transfer: BMP, Excel file
3. Standard Curve
* Input measuring wavelength
* Making a standard Curve: Max. 8 samples
* Selecting Curve: Linear(N), Linear(O), Segmented
* Functional formular /R2 Value indication
* Concentration unit selection
* Graph Zoom In/Out/
* Make a standard curve by inputting Conc. and ABS values.
4. Kinetic mode
* One sample measuring mode
* Measuring time: max. 99hr
* Measuring interval: min. 0.1 sec.
* Changing absorbance/Transmittance data
* Graph Zoom In/Out: max. 100 times
* Confirm/save data as a table
* Merge graph: max. 6
5. ABS Ratio mode

* Setup Numberator/Denominator(260/280nm, DNA, Protein measurement)

* Setup reference wavelength

* Changing measuring wavelength

* Input various factor

6. Custom-Made mode
 
* Residual pesticide analysis(End Point/Kinetics method)

* Soil analysis

* Water quality analysis

* Film thickness/main ingredient analysis(PCB)
Refrigerated BATH CIRCULATOR
Model RC-780
SP-3000plus shown with optional
Sipper system #3000-C-C01
SP-3000plus has
various accessories can be applied various fields.
Accessories can be exchaged easily and conveniently.
Special accessories by user's request can be designed and provided.
Code No. Description  
#3000-C-C03 Micro volume cell holder >300ul, 10mm Pathlength, 8-Position
>30ul, 10mm Pathlength, Single
#3000-C-C08 Film/Thin-layer holder <2mm(Thickness),<10cmx7cm(WxH), Magnetic holder
<5mm(Thickness),<10cmx7cm(WxH), Magnetic holder
#3000-C-C06 Temperature Cell holder single cell holder
Refrigerated Bath Circulator: Temp range:-20¡É~100¡¥ É¥
#3000-C-C04 Long-path cell holder 5cm, Rectangular
10cm, Rectangular
Universal type(Width: 10mm)
#3000-C-C01 Sipper System RS-232C, In take volume700ul
#3000-C-C07 Round cell holder <16mm(D), <100mm(H)
Wavelength Range 190nm~1100nm
Spectral Bandwidth 1.0nm
Wavelength Display Readable to 0.1nm
Wavelength Accuracy ±0.5nm(with automatic wavelength correction)
Data Bunching Interval 1nm
Wavelength Reproducibility ± 0.1nm
Stray Light Less than 0.05% (at 220nm and 340nm)
Photometric Range Absorbance: 00.3~3.0Abs
Photometric System Split-beam Optics, Dual detectors
Photometric Accuracy ±0.002Abs. at 0.5Abs. ±0.005Abs. at 1Abs
Photometric Reproducibility ±0.002Abs at 1.0Abs.
Baseline Correction Automatic
Wavelength Slew Rate About 5000nm/min
Light Source 6V 10W Tungsten halogen lamp and D2 lamp
Monochromator Modified Czerny-turmer type with 1200 limes/mm holographic grating.
Auto Zero Function By a single stroke of Soft key
Display Graphic LCD 6", adjustable viewing angle and contrast control
Detector silicon photo-diode
Sample Compartment Automatically rotative 8 position cell holder
Interface Ports RS-232C and Centronics ports
Power Requirements 220-240VAC, 50/60Hz
Dimensions 514(W) x 243(H) x 430(D) mm
Net Weight 16.0kg
* Continuing research and improvements may result in specification change at any time.
* CE and FDA Listed. Copies of CE and FDA documentation are available for compliance verification

3000-S-V01 OptiView PC Software 30001 Printer(HP Deskjet)

3000-C-C01 Sipper Assembly

30002 Standart Glass Cuvette
3000-C-C03 Micro Volume Cell Holder 30003 500 Micro W/Black Glass Cuvette
3000-C-C04 Long Path Cell Holder 30004 1000 Micro W/Black Glass Cuvette
3000-C-C06 Temperature Cell Holder 30005 Standard Quartz Cuvette
3000-C-C07 Round Cell (dia.16mm) Holder 30006 500 Micro Quartz Cuvette
3000-C-C08 Film/Thin-Layer Holder 30007 1000 Micro Quartz Cuvette
RC-780 R. Bath Circulator(8l, -20~100C) 30008 Didymium Filter
  30009 Printer Cable
  30000 AC Adapter for Printer
 